Easy Stirling Engine Manuals

. SU 1 U -86˚C Ultra Low Storage Operating Manual 03.2012.1 6000 Poston Rd Athens, Ohio 45701, USA 740.274.7900 / 855.274.7900 740.274.7901 www.stirlingultracold.com. ˚ C 03.2012.1 Operating Manual: Ultra Low Storage.

PDF Book quick and easy stirling engine contains important information and. Stirling engine document other than just manuals as we also make available many.

SU105U Quick Guide Initial Setup 1. Unpack freezer unit and included components, ensure all packaging material has been removed from around the fan inlet and outlet holes. Plug AC power cord into the main power socket at the back of the unit and then into an electrical outlet. ˚ C 03.2012.1 Operating Manual: Ultra Low Storage. FeaTURes OF The sU105U FReezeR.8-11 Free-Piston Stirling Engine (FPSE).8 Cabinet Design Features. M150A free-piston Stirling engine (FPSE) technology to provide high effi- ciency, deep-temperature cooling.

1.2 – Applications The SU105U freezer is designed for long-term storage of materials with ultra low, well-regulated temperature storage requirements, such as found in clinical and bio- logical laboratory and testing settings. Please note that precautions and safety instructions are part of each section of the manual, and must be observed to avoid damage to the freezer or harm to its users. Two Model SU105U cabinets stacked. Shown with optional stacking adapter. Two resistance temperature detectors (RTDs) measure the internal cabinet tempera- ture of the SU105U.

One provides input to the FPSE controller and the other provides input for the Swift GUI display and for reporting and logging. The display RTD may be user calibrated through the Swift GUI. Display temperature graph and event log. Display diagnostic and service information. Display freezer identification information Values for the following SU105U parameters are viewed and modified through the touch screen. Modifiable Values Accessible Values Temperature set point Current freezer interior temperature. Recessed Electrical Panel A.

AC Power Connector B. AC Power Switch C. External Alarm Terminal Strip, Dry Contacts (NO, NC, C) D. Battery Power Switch Model SU105U. Shown with optional stainless steel inventory racks.

˚ C 03.2012.1 Operating Manual: Ultra Low Storage. 2.7 – Options. Stacking Adapter When stacking two Stirling Ultracold SU105U FPSE Ultra Low Temperature (-86°C) Freezers, a stacking adapter is required. The Stacking Adapter consists of two (2) plates and twelve (12) screws. 2.7.1 – Adapter Installation Procedure 1. Safety Precautions Potential hazards associated with use of the SU105U freezer may impact the safety of persons in the workplace in which the freezer is installed. This includes persons who operate the freezer, and other personnel in its vicinity.

The freezer itself may also be damaged or its warranty voided by improper operation or usage. Safety Precautions. When removing the power cable, hold onto the plug and not the cable. Do not block the air intake or air discharge vents used by the FPSE (See pictorial on Page 10 for locations.). Ultra low temperature freezers are different from home freezers in that at -86°C frostbite can occur instantly. Unpacking and Installation Verify the SU105U freezer shipment by uncrating the freezer and included items, inspecting them for damage during shipping, and compare the items to the packing list for completeness.

4.1 – Qualify the installation site for the freezer before placing it into service. Unpacking and Installation 4.2 – Set-up 1. Before switching on electrical power to the freezer, make all desired external connections: AC power connection External alarm 2. Turn on the AC switch. If the DC power option has been provided please see the supplemental instructions. Operation 5.1 – Operation of the SU105U freezer comprises four stages: 1. Monitoring temperature performance using the Swift GUI and touch screen control panel.

Managing the freezer contents. Cleaning and defrosting the freezer. Calibrating the freezer's display RTD. Operation 5.1.1 – Swift touch screen control panel: A Password Identification Number (PIN) may be set, but is not displayed.

The default PIN is 1234 and may be changed by the user to any four-digit numeric. Use the PIN to control access to parameter change screens. 5.1.2 – Control Panel Quick Guide The Swift GUI menu tree below starts with the Home screen.

(How to get there is described later.) Soft buttons provide navigation to the screens on the next level of the menu tree. Home – display current freezer temperature. Operation Basic Controls. Generally, to change a value on a screen, Tap the appropriate soft button to highlight the parameter. Use the +/- buttons to enter the desired value and then use the “Save” soft button to complete the change. 5.1.3 – Swift Graphical User Interface (GUI) Instructions 1. Start Screen: Turn On Freezer and Battery Power.

Easy

Stirling Ultracold splash screen will appear for 3 seconds 2. Home Screen: The Home screen will be displayed next Tap the Setup button to proceed to system settings. Data Setup: From the Setup Screen. Tap the Set-up button to change alarm settings, network configurations, date & time, door defrost heater, and PIN. Tap the Silence button to temporarily silence the audible alarm. Tap the Event Log button to track the last approximately 200 events.

Secure Temperature Set Point: To observe the current temperature set point. Tap Change to change the temperature set point and over / under temperature thresholds. Tap the Alarms button to change the alarm delays. Tap the Net button to change network configurations. Alarm Delay: From the Setup Alarms Screen Select the appropriate parameter then use the + and – buttons to adjust then tap Save. (Note: If Save is not tapped new values will not be used).Temp.

Delay delays an over temp alarm after a door opening.Audible Delay delays beeper for a period of time when silence is tapped.Ext. Door Defrost Heater: When set to Auto Defrost, the door defrost heater will cycle 15 minutes ON and 20 minutes OFF unless otherwise programmed. Auto Defrost parameters are not adjustable. To enable Manual Defrost tap the Auto Manual button to activate Manual mode. Operation 11. Event Log: The Event Log will display the last approximately 200.

events in about 30 pages, listing 7 events per page. The last 7 events are shown by default.

Events beyond 30 pages will not be saved. Chart History, Zoom: The Chart History can be enlarged to use the entire display. Tap the screen to enlarge or reduce the Chart History Screen 14. About Freezer: To observe specific freezer information including Model and Serial Num- ber, Firmware Version, Engine Number, Run Hours, Set Point, and Date.

Operation 15. Performance Snapshot: The Service Screen gives you access to system information and calibration, as well as Firmware Upgrade options, Alarm Tests, Diagnostics, Service Contact, and RTD Calibration. Tap the Firmw. Upgrade button to for system firmware upgrades. Diagnostics: Diagnostic Screen provides system health information.

Tap the Back button to return to the Setup Screen. Tap the Home button to take you back to the Home Screen 18. Service Contact: Service Contact Screen provides phone, email, and web address for distributor.

Engine

Operation 19. Calibration: RTD Calibration Screen is used to compensate for discrepancies between the Display RTD and a user provided standard. Tap RTD Select until the desired RTD is selected then use the + and – buttons to adjust the offset. Tap the Calib. Button to save changes.

5.3 – Cleaning Cleaning the Freezer Surfaces. Clean the exterior surfaces of the SU105U cabinet as needed by using a soft cloth and mild detergent.

Do not use solvent (such as bleach) or harsh abrasive cleansers or pads. Contents, Cleaning, Transportation 5.4 – Calibration of the RTDs The SU105U Display RTD and Control RTD may be calibrated by inserting a temperature sensor through the access port of the cabinet wall and securing it next to the RTDs located in the right rear corner of the interior just above the shelf. Power Consumption 350W maximum Maximum Current 3.0 A at 110 VAC / 1.5 A and 240 VAC Cooling System / Direct cooling by a Free Piston Stirling cooler Device and thermosiphon heat transfer Set Temperature Range -86°C to -20°C Ambient Operating +5°C to +35°C (at RH 65%). 7.

Specifications 7.2 – Materials Specifications Part Material Color Treatment Body: Main Body, Door, Mild steel White Powder coated Front Cover and Top Cover Interior Stainless Steel Stainless Steel Shelf (non-adjustable) Handle Aluminum Gray Powder coated Hinges Steel White Powder coated LCD Bezel Polycarbonate Gray. Troubleshooting Problem Possible Cause Solution Check the connections to the power supply Freezer does not Power is not getting to the power on freezer from the Power Supply Assure that the wall plug is firmly seated Inadequate air circulation Remove airflow obstructions Assure that the wall plug is Inadequate power. Troubleshooting Note This freezer is designed for storage of frozen product.

Stirling Engine Generator

For best results:. Always operate your freezer at full capacity with real or simulated product to increase the thermal mass, displace air and maintain optimal stability.

Always minimize the frequency and duration of door openings. The warranty period starts TWO WEEKS after the original date of shipment from Global Cooling, Inc.

Stirling Engine For Sale

The Stirling Ultracold freezer is warranted for a period of TWO YEARS for materials and labor. The Stirling Ultracold Humm™ free-piston Stirling engine and thermosiphon is war- ranted for a full SEVEN YEARS, parts only, from original date of shipment from Global Cooling, Inc. Global Cooling further warrants all products to be free from defects in materials under normal use and service for a period of TWO YEARS. The Stirling Ultracold Humm™ free-piston Stirling engine and thermosiphon is war- ranted for a full SEVEN YEARS, parts only, from original date of shipment from Global Cooling, Inc. ˚ C 03.2012.1 Operating Manual: Ultra Low Storage. CE Documentation 10. CE Documentation t: 855.274.7900 www.stirlingultracold.com.

©2012 Stirling Ultracold, Global Cooling, Inc. All Rights Reserved. HUMM™ technology is manufactured under U.S. And International patents. Stirling Ultracold, Shuttle™, HUMM™ and Swift™ are trademarks of Global Cooling, Inc. Specifications subject to change without notice. Refer to www.stirlingultracold.com for the latest specifications.